Linksens focused on high-end intelligent water quality monitoring equipments and solutions
The S‑RC051 is a three‑electrode constant potential current measurement sensor developed by our company with independent intellectual property rights.
It can be used to measure the concentration of disinfectants such as residual chlorine, high‑purity gas dioxide, and ozone. The electrodes incorporate an ARM processor and a high‑efficiency filtering algorithm to effectively avoid noise interference
Water supply pipelines;Tap water at the outlet of waterworks;Terminal water of water supply pipelines;Secondary water supply;Swimming pools.
| Digital Disinfectant Residual Chlorine Sensor | |||||
| System | |||||
| Measurement Parameters | Residual chlorine,chlorine dioxide(high purity) Ozone,total chlorine,etc | ||||
| Water Sample Flow Rate | > 400 mL/min,and the flow rate shall be kept stable | ||||
| Working Voltage | DC9 - 28V | ||||
| Peak Power | 0.5W | ||||
| Digital Output | RS485 Modbus protocol (optional4 - 20mA) | ||||
| Electrode Material | Platinum/Gold | ||||
| Disinfectant | |||||
| Measurement Range | 0 - 5/20mg/L | ||||
| Accuracy | ±3% (DPD comparison error ±10% or ±0.05 mg/L, whichever is greater) | ||||
| Response Time (T90) | ≤90s | ||||
| Measurement Method | Reagent - free, electrochemical three - electrode amperometric assay system | ||||
| Repeatability | ±3% | ||||
| Calibration Method | Single - point calibration, Two - point calibration | ||||
| Resolution | 0.001mg/L | ||||
| Measurement Lower Limit | Lower detection limit: 0.030 mg/L or less | ||||
| Sample pH Range | 4 - 9 | ||||
| Temperature | |||||
| Measurement Range | - 5 - 60℃ | ||||
| Measurement Method | Thermistor method | ||||
| Accuracy | ±0.4℃ | ||||
| Resolution | 0.1℃ | ||||
| Calibration Method | Single point calibration | ||||
| Repeatability | 0.2℃ | ||||
Main Features